《丹麦GRAS40BP麦克风技术规格指南.docx》由会员分享,可在线阅读,更多相关《丹麦GRAS40BP麦克风技术规格指南.docx(3页珍藏版)》请在优知文库上搜索。
1、丹麦GRAS40BP麦克风技术规格指南频率范围:4Hz至70kHz动态范围:34dB(八)至169dB灵敏度:1.6mV/PaFrequencyrange(1dB)Hz10to25kFrequencyrange(2dB)Hz4to70kDynamicrangelowerlimit(microphonethermalnoise)dB(八)34DynamicrangelowerlimitwithGRASpreamplifierdB(八)39DynamicrangeupperlimitwithGRASpreamplifier+120V/260VpowersupplydB169Open-circuit
2、sensitivity250Hz12dB)mV/Pa1.6Open-circuitsensitivity250Hz(2dB)dBre1VPa-56ResonancefrequencykHz50Microphonecartridgecapacitance,typ.pF7MicrophoneventingRearTemperaturerange,operationCF-40to150/-40to302IEC61094-4DesignationWS3PTemperaturerange,storageC*F-40to85/-40to185Humidityrangenoncondensing%RHOto
3、90Temperaturecoefficient250HzdBC/dBF-0.01/-0.006Humiditycoefficient250Hzd%RH-0.0013Staticpressurecoefficient250HzdkPa-0.008Influenceofaxialvibration1msadBre20Pa55CE/RoHScompliant/WEEEregisteredYes/Yes.YesWeightg/oz1.5/0.053PolarizationZConnection200V/Traditional40BP是IEC61094WS3P1/4外极化压力麦克风,带后通风。它是根据
4、IEC61094-4要求制造的高精度电容式麦克风。它是一种低灵敏度麦克风,用于高电平和高频的声音测量。其低灵敏度使其非常适合测量高达174dB的高声压级。它的小尺寸减少了麦克风周围的衍射和反射影响,从而使频率范围扩展到70kHz040BP在出厂时经过单独校准,并随附校准图表,说明其特定的开路灵敏度和压力频率响应。典型应用和用途40BP通常用于声压、高频和高电平压力测量。它适用于耦合器和边界处的通用声学测量。40BP是一个非常通用的麦克风,当连接到RA0086发射器适配器用于1/4麦克风时。当RA0086直接从信号发生器获取校准信号并使麦克风表现得像静电时,40BP成为一个高阻抗声源扬声器。当用
5、于校准声耦合器时,这会提供与40BP表现得像麦克风时一样好的频率响应。注意:在低频时,保护网格在声音上是透明的;然而,在20kHz及以上,保护网格可以充当谐振器并影响频率响应。因此,对于更高频率的测量,英寸和%英寸麦克风已经过优化,可以在没有保护栅的情况下使用。测量时请取下保护栅。兼容性40BP需要标准化的1/2或1/4LEMO前置放大器和支持该技术的输入模块,带有7针LEMO连接器。系统验证对于测量设置的口常验证和检查,我们建议使用校准器,如GRAS声级校准器42AGo为了正确校准灵敏度,我们建议使用活塞发声器,如GRAS智能活塞发声器42APo校准出厂时,所有GRAS麦克风均已在受控实验室
6、环境中使用可追溯校准设备进行校准。根据使用情况、测量环境和内部质量控制程序,我们建议麦克风至少每年重新校准一次。我们提供两种校准作为可选的售后服务:GRAS可追溯校准和GRAS认可校准。GRAS可追溯校准是由经过培训的人员在受控条件下根据既定程序和标准执行的可追溯校准。这与作为我们质量保证不可或缺的一部分的所有GRAS麦克风都要进行的严格校准相同。GRAS认证校准由GRAS认证校准实验室执行,该实验室已根据ISO17025由丹麦认证基金DANAK进行认证。所有GRAS话筒均由优质材料制成,可确保终生的稳定性和坚固性。麦克风全部由在该领域具有多年专业知识的熟练且敬业的操作员在经过验证的洁净室环境中组装。麦克风振膜、主体和改进的保护栅均由高级不锈钢制成,这使麦克风能够抵抗物理损坏以及腐蚀性空气或气体引起的腐蚀。